2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 607
P- F.7.0 ST 12 LESS THAN 23.0 VOLTS/ ISOLATE MODULES / PUMPS
a)
IMPORTANT! To prevent damage, turn the machine OFF!
b) Figure below, unplug the ACTUATOR Cable from the distribution board.
c)
ENSURE the meter’s black lead remains connected to chassis ground!
d)
IMPORTANT! Turn the machine ON (fan running)!
e) Figure below, measure again from the rear side of the Power Control Board’s ST12 connector. More
than 23 volts DC now?
Yes
More than 23.0 volts! Possible bad Acid, Bicarbonate, Flow or Deaeration Pump. Check
their distribution board connectors and wire harnesses for damage.
No
Less than 23.0 volts! See procedure number