2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 621
P- H.1.5 SYMPTOM STILL PRESENT
a)
To avoid damage, turn the machine off!
b) Return the Functional and Power Logic board device cables to EXACTLY as recorded above!
c) Referring to see Figure 4A and NOTE A (page 9), if (and ONLY if) the machine is equipped with the
New Style Power Logic Board (i.e. NO ‘Converter Board’) skip to procedure number
622). If the ‘Converter Board’ is present see procedure number
(page 621).
P- H.1.6 ISOLATE CONVERTER BOARD (IF PRESENT)
a) CAREFULLY remove the Converter Board from the motherboard’s nine (9) pin TEST connector.
b) If the screen is NOT turning ON skip to part d. If the screen IS turning on enter
Service Mode → Options → Hardware Options. Next to
T and C Mode
place the
‘X’ in the “Yes” box and press the ‘CONFIRM’ key (the ‘X’ turns blue).
c) Turn the machine ON (fan runs).
d) If possible, place it into Dialysis or RINSE.
IGNORE the ‘grinding sound’ and all hydraulic alarms!
e) ALLOWING sufficient time for the symptom to develop, either measure for +12 and/or +5 volts DC
(V
DC
)
,,
OR call debug screen 1 to monitor
5V Est
and/or
12V Est
. Within range and stable now?
Yes
Good voltages and stable! The Converter Board may be causing the problem!
CAUTION!
The new Convertor Board MUST be installed with the ‘solder side’ towards the front of
the machine AND matched pin for pin to the TEST connector!
(page 622).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...