2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 478
PHT- 2.0.0 START PRESSURE TESTS
a) RESET ALL alarms!
b) At the bottom of the screen, select the
[Test & Options]
tab. TWO (2) possible scenarios:
1)
IF the [Pressure Test] button is BLUE:
Select
[Pressure Test]
THEN press ‘CONFIRM’ then
IMMEDIATELY continue to procedure number
(page 478).
2) IF the [Pressure Test] button is GRAY:
Select
[Both Tests]
THEN press ‘CONFIRM’ then
QUICKLY read parts a AND b below:
a) Both Tests start with “Get Neg TMP”. If (and ONLY if) this initial “Get Neg TMP” period lasts
longer than two (2) minutes seconds see ABOVE procedure number
(page 475). If
“Get Neg TMP” lasts less than two (2) minutes continue to part b.
b) ENSURE all tests above Negative Pressure Tests pass
*.
When “Conductivity Low Soft”
appears IMMEDIATELY continue to procedure number
(page 478).
*
Other than Battery all other test failures above Negative Pressure may cause the Pressure
Tests to fail.
PHT– 2.0.1 PRESSURE TESTS RUNNING / ANALYZE TMP
a) Call debug screen 1.
b) The Pressure Tests start with “Remove Air” (
FILACT
= 1) followed
by “Get Neg TMP”. The UF Pump strokes every eight (8) seconds
until
ATMP
(Figure right) reaches about -250 mmHg then “Neg Flow
On” appears (for one minute) followed by “Pos Flow Off” (for one
minute).
c) Is
“Get Neg TMP”
lasting longer than ninety (90) seconds?
Yes
“Get Neg TMP” longer than 90 seconds! See procedure number
(page 478).
No
“Neg Flow On” appeared! Proceed to
page 480
, procedure number
PHT- 2.0.2 “GET NEG TMP” LASTING LONGER THAN 90 SECONDS
Watch
FILACT
for one (1) minute while ALSO ENSURING “Get Neg TMP” REMAINS up.
Does
FILACT
EVER = 1, even if just once?
Yes
FILACT =
1 even if only once WHILE “Get Neg TMP” is up! Possible air leak. Proceed to
page
SECTION 8 – FILLING PROGRAM PROBLEMS
No
FILACT
ALWAYS = 0! See procedure number
(page 479).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...