2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 480
PHT- 3.0.0 “NEG FLOW ON” APPEARED
Allow the tests to finish! Do BOTH Pressure Tests (Positive and Negative) pass?
Yes
BOTH pass!
IMPORTANT!
Proceed to
page 484
, procedure number
No “Pressure Test Failed”! At the bottom of the screen, select the [Test & Options] tab. TWO (2)
possible scenarios:
1)
IF “Fail: Remove Air” is listed:
NOTE it THEN see continue to procedure number
2) ALL OTHER failure listings:
PHT- 3.1.0 PRESSURE TEST(S) FAILED
a) Call debug screen 1 to locate
ATMP
and
FILACT
(Figure right). Below
you will watch these while ALSO watching the screen banners!
b) Reset ALL alarms!
c) Select the [Test & Options] tab THEN the [Pressure Test] button.
d) Press ‘CONFIRM’ THEN
call debug screen 1.
e) During the sixteen (16) second “Remove Air” period
FILACT
= 1.
f) When “Neg Flow On” first appears a thirty (30) second STABILIZATION period begins.
ATMP
should
settle between -250 and -440!
NOTE:
If “Remove Air” reappears the tests are REPEATING AND
FILACT
will = 1 again. In this
event, allow “Neg Flow On” to reappear i.e. the repeat STABILIZATION period.
NOTE:
If
FILACT
EVER = 1 during the repeat STABILIZATION period NOTE this!
NOTE:
If
ATMP
does not remain between -250 and -440 “Fail Neg Stabilize” will be listed in the Test
& Options screen possibly indicating a large air leak.
g) If
ATMP
remains
between -250 and -440 a thirty (30) second HOLD period begins. “Neg Flow On”
remains up AND
ATMP
should not change more than +/- 20!
NOTE:
If
ATMP
increases to more than -440 NOTE this!
NOTE:
If “Remove Air” reappears the tests are REPEATING AND
FILACT
will = 1 again! In this
event, allow “Neg Flow On” to reappear then thirty (30) seconds for the HOLD period to begin
again.
NOTE:
If
FILACT
EVER = 1 during the repeat HOLD period NOTE this!
NOTE:
If
ATMP
does not HOLD +/- 21 “Fail Neg Flow On” will be listed in the Test & Options screen
possibly indicating a small air leak.
Parts h through k next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...