2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 484
PHT- 4.0.0 BOTH PRESSURE TESTS PASS
a)
ENSURE
the Loading Pressure gauge (yellow connector) reads 0 psi before inserting it!
b)
SLAM
the gauge into the red Acetate/Acid rinse port.
*
ELSE pressure will not be read correctly
c) Loading Pressure is ‘OKAY’ if it cycles between a peak of somewhere between 23 and 25 psi AND
NEVER lower than 11 psi at its lowest.
d) Watch the gauge for one (1) FULL minute. FOUR (4) possible scenarios 1) or 2) or 3) or 4) below:
1)
IF (and ONLY if) Loading Pressure REMAINS ‘OKAY’:
2)
IF (and ONLY if) peaking to MORE THAN: 26 psi:
Turn Valve #65’s nut (Figure below)
counterclockwise (outward). If a PEAK pressure between 23 and 25 psi CAN be achieved see
procedure number
(page 485). If (and ONLY if) NOT Valve #65 may be bad.
3)
IF (and ONLY if) NEVER less than 11 and peaking to between 15 and 22 psi:
Adjust Valve
#65’s nut (Figure below), clockwise until a peak of between 23 and 25 can be achieved! If good
pressure CAN be adjusted see procedure number
(page 485). If (and ONLY if) it
CANNOT proceed to
page 27
procedure number
F- 1.0.8
.
4)
ALL other scenarios:
ENSURING the gauge was SLAMMED into the Rinse Port proceed to
page 27
procedure number
F- 1.0.8
.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...