2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 242
T- 1.3.70 RETURN SYSTEMS / TEMP COMP?
a)
IMPORTANT!
Plug
NTC #2’s connector into the
1
st
distribution board position from the left, “CON-NTC”!
b)
IMPORTANT!
Avoiding VACANT position #4, RETURN
NTC #3’s connector to the
2
nd
distribution board position
from the left, “MON-NTC”.
c)
Enter Service Mode → Options → Hardware Options.
Does the “Yes” box at
‘Temp Comp’
have a blue ‘X’ in it?
(Yes or No)
Yes
Place the ‘X’ in the “No” box and press ‘CONFIRM’. The ‘X’ turns blue!
Post a note that ‘
Temp Comp
’ is off then see procedure number
(page 242).
No Temp Comp = No! See procedure number
T- 1.3.80 ISOLATE TEMPERATURE INTERNAL LIMITS
a) Select C
alibrate Sensors → Temp Control.
b) Connect a temperature meter to the dialyzer connectors.
c) Sharply press ‘CONFIRM’ TWICE to turn the
[TEMP DAC]
data box pale yellow / white.
d) Select
[TEMP DAC]
, it turns bright yellow.
e) Set
[TEMP DAC]
to “130” and press ‘CONFIRM’. ENSURE the
[TEMP DAC]
is pale yellow / white. If
gray exit the calibration then return to part a!
f)
IMPORTANT!
Allow five (5) minutes BEFORE continuing to part g!
g) Press ‘CONFIRM’ twice to activate the
[Monitor Reference]
data box. It turns bright yellow.
h) IGNORING the external meter adjust the
[Monitor Reference]
data box
to 34.0° C!
i)
Press ‘CONFIRM’ twice to save the calibration.
j)
(page 243).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...