2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 180
CLEAN- 4.6.0 VERR = 0 / ISOLATE DIALYSATE PRESSURE SENSOR
When
Remaining Prerinse Time
= 0:00, Valve #29 stays open and Valve #30 stays closed. The Actuator-
Test Board watches for Dialysate Pressure Sensor #9 (debug screen 4’s
PDIA
and
ADIA
) transitions to
sense flow.
a) Call the “Interrupted” banner (Figure right) by pressing
‘Escape’ TWICE then ‘CONFIRM’ ONCE!
b) ENSURING the “Interrupted” banner is up, remove the red DIALYZER connector from the shunt door
and hold it at the level of the dialyzer holder. ENSURE no flow from dialyzer connector!
c) Call debug screen 4 (Figure right).
d) Locate
PDIA
then its value in
Column 1 of the Table below.
e) Locate
ADIA
then its value in
Column 2 of the Table next to the
value seen at
PDIA
.
f) Per Column 3 respond based on
BOTH
PDIA
AND
ADIA
:
Column 1
PDIA
Column 2
ADIA
Column 3
Your Response
Between 3.0 and 6.0
(Good)
Between 2.0 and 6.0
(Good)
BOTH
PDIA
and
ADIA
are good! The Dialysate
Pressure sensor is good! See procedure number
(page 181).
Between 3.0 and 6.0
(Good)
Less than
2.0 OR more than 6.0
(Bad)
PDIA
good;
ADIA
bad. THREE (3) possibilities:
1) Bad Actuator-Test Board OR
2) Bad Sensor Board OR
3) Bad Motherboard connection at the Sensor
and/or the Actuator-Test Board
Less than
3.0 OR more than 6.0
(Bad)
Between 2.0 and 6.0
(Good)
PDIA
bad;
ADIA
good! THREE (3) possibilities:
1) Bad Sensor Board OR
2) Bad Functional Board OR
3) Bad Motherboard
Less than
3.0 OR more than 6.0
(Bad)
Less than
2.0 OR more than 6.0
(Bad)
BOTH
PDIA
AND
ADIA
are
bad! Leaving the
“Interrupted” banner up, proceed to
page 462
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...