2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 299
T- 6.2.0 ISOLATE AIR REMOVAL SYSTEM
a) Turn the water off.
b) The Deaeration gauge is used next.
ENSURE
it reads 0
before installing it!
c) After the No Water” alarm appears allow one (1) minute
for the Deaeration motor to stop.
d) Figure right, tee the Deaeration gauge into the Input
(clear tubing) side of Deaeration Pump #20.
e) Turn the water on and allow the “No Water” alarm to go
away.
f) Is Deaeration Pressure OKAY? Refer to
(page 667) for what pressure should be.
Yes
Deaeration pressure is okay. Position the gauge so that its tubing is not pinched THEN see
procedure number
No Deaeration pressure is NOT OKAY! ENSURING the “No Water” alarm is NOT presenting,
NOTE this page number, as you will return here, THEN proceed to
page 498
T- 6.3.0 ISOLATE POTENTIAL AIR LEAKS
Per the Figure right, using a flashlight, WITHOUT LOOKING
AWAY for two (2) minutes watch for air bubbles through the
tubing at Conductivity Cell #7!. Air seen?
Yes
Air seen! Proceed to
page 495
, procedure
No air seen! See procedure number
(page 299).
T- 6.5.0 ISOLATE BALANCING CHAMBER DIAPHRAGM
Was [
Temperature
] EVER
observed ‘bouncing wildly’ up for
½ second then returning to normal
(Yes or No)?
Yes
Another procedure in different Section is performed next.
IMPORTANT! NOTE
this page and
procedure number because you may prompted to return to here. Perform parts a and b below
a) BEFORE continuing to part b, proceed to
page 535
, to perform
FOR A LEAKING BALANCING CHAMBER DIAPHRAGM
b) If a torn balancing chamber diaphragm was not located in part a see procedure number
(page 300).
(page 300).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...