2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 623
P- H.3.0 ISOLATE 24 VOLT CONNECTIONS
a)
TURN THE MACHINE OFF!
b) Open the power supply enough to see
the top edge of the Power Control
Board (Figure right).
c) ENSURE the 24V Power Harness is
connected SECURELY, pin for pin, to
its connector!
(page
P- H.3.1 SYSTEM INITIALIZATION?
a) Turn the machine on (fan running).
b) Allowing forty (40) seconds does System Initialization reach 100%?
Yes
System Initialization reaches 100%! See procedure number
No System Initialization does NOT reach 100%! Proceed to
page 638
, procedure number
P- H.3.2 SYSTEM INITIALIZATION REACHES 100%
NOTE:
The previous Voltage Detection calibration may have solved the problem!
NOTE:
If the 24V Power Harness was not securely connected this may have been the problem!
a) Place the machine into Dialysis OR Heat Disinfect OR Rinse.
b) Does a voltage OR “Failed Sending Data” OR “Temp Over 95 Degrees” alarm occur?
Yes
Alarm occurs! See procedure number
(page 623)
No
An alarm does not occur! Allow thirty (30) minutes or longer. If a voltage OR “Failed Sending
Data” OR “Temp Over 95 Degrees” alarm reoccurs see procedure
(page 623). If these
alarms DO NOT reoccur the troubleshooting Guide cannot locate an immediate problem.
P- H.3.3 ALARM REOCCURS
a) Turn the machine off then back on HOWEVER, when ”System Initialization” reaches 100%
DO NOT
press any keys yet!
b) Per the OBSERVED symptom, FOUR (4) possible scenarios 1) or 2) or 3) below:
1)
IF “5V High” OR “5V Low” OR “12V Power Fail”:
(page 624).
Scenarios 2 through 4 next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...