2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 163
CLEAN- 1.2.6 MOTOR CONTINUES TO ROTATE
Recheck Loading Pressure (gauge in Rinse port). THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) ALWAYS less than 15 psi:
Leaving the machine in the Cleaning Program, see
(page 163).
2)
IF (and ONLY if) MORE than 25 psi:
Figure below,
turn Valve #65’s nut counterclockwise (outward) to
adjust to between 23 and 25 psi then proceed
page 172
, procedure number
3)
IF cycling to between 15 and 21 psi:
Figure below, turn Valve #65’s nut clockwise (inward)
attempting to adjust pressure to a PEAK between 23 and 25 psi. Can you?
Yes
Pressure CAN be adjusted to between 23 and 25 psi! Allow two (2) minutes then if a Flow
Error reoccurs proceed
page 172
, procedure number
No
Loading Pressure remains low! Leaving the machine in the Cleaning Program, see
procedure number
(page 163).
CLEAN- 1.5.0 LOADING PRESSURE REMAINS LOW
a) Turn the Heater Breaker Switch OFF!
b) Figure right, unplug the Deaeration Pump from
distribution board position, “P20, DEGAS-P”.
c) ENSURE the Deaeration Motor has stopped
rotating!
The pump remains unplugged
till instructed!
d) Obtain a
1000 ml
graduated cylinder and a bucket.
(page 164).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...