2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 94
F- 9.0.24 ISOLATE FLOW PUMP INPUT FLOW
a) Obtain a 1000 ml (or larger) graduated cylinder!
b)
Service Mode’s Flow Pressure is used to isolate the Flow Pump’s input circuit!
DO NOT follow the screen instructions! Perform no adjustments until instructed!
c)
Enter Service Mode → Calibrate Hydraulics → Flow Pressure.
d) Press ‘CONFIRM’ THEN ENSURE the FLOW MOTOR
shaft is rotating!
e)
NOTE
the gauge reading for later.
f) Figure right, remove the Flow Pump’s INPUT (clear)
tubing and measure flow from it for one (1) minute.
g) Press CONFIRM to return to “Calibrate Hydraulics”.
h) Reattach the Flow Pump’s tubing!
i)
TWO (2) possible scenarios:
1)
IF (and ONLY if) MORE THAN 800 ml collected!
(page 94).
2) IF LESS THAN 800 ml collected!
Proceed to
page 97
, procedure number
F- 9.0.25 INPUT FLOW PUMP MORE THAN 800 ML
a)
Remove the clamps from Valve #43!
b) As noted above, was pressure between 35 and 36 psi?
Yes
Between 35 and 36 psi! Good static pressure but NOT dynamic! TWO (2) possible bad
components:
1)
Flow Pump head OR;
2)
Flow Pump motor (possibly brushes).
No Less than 35 psi! See procedure number
(page 95).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...