
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 40
F- 3.0.0 ISOLATE VALVE #43 (FILLING PROGRAM) DRAIN FLOW
A) WITHOUT returning the Air Sensor’s connector yet, remove the resistor plug from the distribution board.
B) ENSURE debug screen 0’s Chamber #69 says “Air”.
C) Obtain a 1000 ml graduated cylinder!
D) Figure right, ENSURE a “Dial Valve Failure” OR “Act
Byp Valve Fail” banner NEVER appears!
E)
Call debug screen 1. FILACT MUST = 1 before continuing!
(page 40).
F- 3.1.0 FILACT = 1 / ISOLATE VALVE #43 FLOW
a) Figure right, if a ‘Quick Connector’ is present at
the end of the drain tubing an
adaptor
is
required!
b) Measure drain flow for two (2) minutes. 600 ml or
more collected?
Yes
600 ml or more! Reconnect the drain then proceed to
page 42
, procedure number
No Less than 600 ml! See procedure number
(page 41).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...