2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 409
SECTION 8 - FILLING PROGRAM (AIR LEAK) PROBLEMS
Filling Programs NORMALLY occur when air is sensed by Chamber #69’s Air
Sensor #6 i.e. Chamber #69’s text box says “Air”. They are indicated with a “Filling
Program” or “10 Fill Pgm in 1 Hr” or “TMP Is Low” banner. Also, debug screen 1’s
FILACT
= 1.
A) ENSURE the dialyzer connectors are attached PROPERLY to shunt door!
B) Close the shunt door!
C) ENSURE the external flow indicator’s (sight tube) fittings are tight!
D) From the Home screen, ENSURE
[Dialysate Flow] has been on and set to 500 ml/min for six (6)
minutes.
E) The Treatment Clock (Figure right) must be left off i.e. “Tx Paused”.
F) USING A FLASHLIGHT, ENSURE no air bubbles MOVING into the machine
through the Acid and Bicarb inlet tubing!
G) [
Conductivity
] MUST BE more than 13.0 mS BEFORE continuing to part H!
H) If the Automated Tests are running (screen reads “Test:.…”) allow them to finish.
I)
Remove the ‘dummy venous chamber’ from the Level Detector, till instructed!
J) Select the [Dialysate Flow] window (it turns bright yellow).
K) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
L)
DO NOT reset alarms!
M) Is the external flow indicator’s ‘bob’ rising at least ¼ way up in the sight tube?
Yes
‘Bob’ moving! See procedure number
(page 410).
No a) Call debug screen 0.
b) WITHOUT LOOKING AWAY, watch
Flow Error
for two (2) minutes. If EVER = 1, even if only
once, indicates a Flow Error! TWO (2) possible
scenarios 1) or 2) below:
1)
IF (and ONLY if) Flow Error REMAINS ALWAYS = 0!
Call the Home screen.
[
Temperature
] MUST BE between 35.1 and 38.9° C; [
Conductivity
] more than 13.0
mS BEFORE continuing to procedure number
(page 410).
2) IF Flow Error EVER = 1!
Proceed to
page 20
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...