2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 72
Figure 16 – Valve #25
F- 7.0.6 ISOLATE VALVE #26 INPUT FLOW
(previous page)
, clamp
VALVE #26’s
INPUT tubing.
b) Remove the tubing and route it into the cylinder.
c) Remove the clamp and measure flow, from the tubing, for one (1) minute. 800 ml or more collected?
Yes
800 ml or more!
A)
Clamp and reattach the tubing;
B)
(page 72).
No Less than 800 ml!
A)
Clamp and reattach the tubing;
B)
Proceed to
page 74
, procedure number
F- 7.0.7 ISOLATE VALVE #26 OUTPUT FLOW
a) Figure right, remove the Flow Pump’s INPUT (clear) tubing.
b) Remove the clamp from Valve #26 and measure from the
Flow Pump tubing for one (1) minute. 800 ml or more
collected?
Yes
800 ml or more! Clamp the tubing at VALVE
#26, reattach it, then see procedure number
(page 73).
No Less than 800 ml! See parts a THROUGH c below:
a) Reattach the Flow Pump’s tubing AND return valves #37 AND #30 to their distribution
board positions!
b) Using a flashlight, check for restrictions through the tubing between valve #26 and Air
Removal Chamber #69.
c) Assuming all above procedures were performed correctly, TWO (2) possible bad
components:
1)
Actuator-Test
*
Board OR;
2)
Valve #26.
*
To LOCATE the board refer to
(page 9).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...