2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 380
CO- 10.0.20 PROBLEM WITH POST DIALYZER TEMP / ISOLATE TEMP SENSOR #44
a) Figure right, from the distribution board, unplug the
11
th
Connector cap from the left. This is
Temperature Sensor NTC #44, “x44, NTC-POST”!
b) Screen 0’s
Post Dialyzer Temp
MUST be LESS THAN
34.0° C at this time! If NOT, ENSURE Post Dialyzer
Temperature Sensor #44 was unplugged!
c) Using a flashlight, check the vacant “x44” position.
If
corrosion or damage is located this may be the
problem!
d) Figure right, install the
6.04 KΩ
plug, from the
TWO-RESISTOR SET
,
into the Post Dialyzer Temp Sensor’s #44 distribution board position, “x44,
NTC-POST”.
e) Is screen 0’s
Post Dialyzer Temp
between 35.5 and 38.5° C?
Yes
Post Temp between 35.5 and 38.5! Proceed to
page 382
No Post Temp is NOT between 35.5 and 38.5! See parts a and b below:
a) ENSURE the
6.04 KΩ
plug from the
TWO-RESISTOR SET
is placed properly at
position “x44, NTC-POST”. If NOT, repeat (ABOVE) procedure number
380) AND / OR CONSIDER using the
6.04 KΩ
plug from another
TWO-RESISTOR SET
.
b) If SURE the plug is placed correctly see procedure number
CO- 10.0.22 POST DIALYZER TEMP IS NOT BETWEEN 35 AND 38°C
a) ENSURE connector labeled #7 (Pre Cond Cell #7) is in position “X7, COND”.
b)
Enter Service Mode → Calibrate Sensors →
Post Temp Sensor
.
NOTE:
This is a non-routine
troubleshooting procedure! Follow the instructions carefully to avoid error!
c) Press ‘CONFIRM’. The screen says
“1. Connect a 6.808 K ohm
resistor…”
.
d) Per the Figure right, place the
34° C
(6.808 KΩ) plug, from the
FOUR-
RESISTOR SET
into distribution board position, “x44, NTC-POST”.
e) Is the screen’s [
Post-Temperature Reference]
more than 20?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...