2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 147
1)
IF (and ONLY if) ALWAYS LESS THAN 18 psi:
2)
IF the PEAK cycles to or remains MORE THAN 18 psi:
THREE (3) possible scenarios below:
Scenario #1:
IF (and ONLY if) the MINIMUM EVER falls lower than 8 psi
: This indicates a
problem with RO Product flow or pressure!
Scenario #2: IF (and ONLY if) the MINIMUM remains more than 8 psi AND a “No Water”
alarm reoccurs:
Proceed to
page 151
, procedure number
Scenario #3:
IF the MINIMUM remains more than 8 psi AND a “No Water” DOES NOT
reoccur:
The problem is not presenting at this time!
NW- 1.0.14 PRESSURE REMAINS LESS THAN 18 PSI
a) Figure right, loosen Inlet Pressure Regulator #61’s
Lock Nut.
b) Watching the gauge, turn Regulator #61’s
center Adjustment Bolt clockwise (inward)!
TWO (2) possible scenarios:
1)
IF (and ONLY if) MORE THAN 18 psi CAN
BE achieved: A)
Adjust Regulator #61 until
the PEAK is between 18 and 20 psi
;
B)
Return to (ABOVE) procedure number
page 146
2)
IF MORE THAN 18 psi CANNOT be achieved:
ASSUMING incoming water has remained more
than 1000 every minute, Inlet Pressure Regulator #61 may be bad.
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...