2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 452
TMP- 2.0.5 FILACT WAS OR IS = 1
This procedure ENSURES Valve #43 is not sticking closed AND allows
Conductivity
to increase:
a)
IMPORTANT!
From the Home screen, select the [Dialysate Flow] window!
b) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’!
c) Call debug screen 0. WITHOUT LOOKING AWAY, watch
Flow
Error
for four (4) minutes. It should NEVER = 1! TWO (2)
possible scenarios:
1)
IF (and ONLY if) Flow Error EVER = 1 , even if only once:
Proceed to
page 20
1, FLOW ERRORS IN DIALYSIS PROGRAM
2)
IF Flow Error ALWAYS = 0!
See parts a THROUGH e below
a) From the Home screen, allow [
Conductivity]
to stabilize between 13.0 and 14.5 mS!
b)
IMPORTANT! Open the shunt door and LEAVE IT OPEN till instructed!
c)
Install a ‘dummy venous chamber’ in the Level Detector module!
d) Call debug screen 1. Allow
FILACT
= 0 ALWAYS. If
FILACT
DOES NOT = 0, after two (2)
minutes, most likely, the female Air Sensor connector is NOT plugged in PROPERLY!
e)
ALARM RESET SEQUENCE:
i)
Press and release the ‘Reset’ key then immediately press
and hold it for three (3) seconds;
ii)
Allow thirty (30) seconds. If ANY alarm reoccurs attempt
RESET up to twice more BEFORE continuing to procedure number
TMP- 2.0.6 ISOLATE UF PUMP CONTROL / UF PUMP LEAK
a) ENSURE the lamp above the front panel’s UF on/off key is
OFF
i.e. UF is
OFF!
b) Call debug screen 0.
c)
Figure right
,
WITHOUT LOOKING
AWAY, watch the UF
Pump’s ‘dot’ for forty five
(45) seconds. It MUST
REMAIN white! Does it
EVER ‘blink’ to blue (Yes
or No)?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...