2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 610
P- G.5.0 ORIGINAL SYMPTOM STILL PRESENT / ISOLATE MODULES
NOTE:
The previous Power Logic Board is probably good!
a)
IMPORTANT! To prevent damage turn the machine OFF!
b)
Per the Figure previous page
, unplug the
MODULES ONLY
, including the Blood Pressure
module!
c)
IMPORTANT!
Turn the machine ON (fan running)! If possible, return to Dialysis OR Heat Disinfect.
d) Re-m12 and / or +5 volts (within range now?). If the symptom was intermittent call debug
screen 1 to watch
5V Est
and / or
12V Est.
Allowing sufficient time, does the symptom you were
troubleshooting reoccur?
Yes
Symptom reoccurs!
A)
IMPORTANT!
Turn the machine OFF!
B)
BEING CERTAIN
the
Blood Pressure module
cable
ONLY
is returned to the Blood Pressure or Colin connector,
reconnect ALL module cables
; C)
Proceed to
page 611
, procedure number
No
Screen on and / or good voltage(s) now! To locate the bad module see procedure number
(page 610).
P- G.5.1 SYMPTOM NOT PRESENT / ISOLATE POTENTIAL BAD MODULE
a)
IMPORTANT!
To prevent damage turn the machine OFF!
b) Your choice, plug ONE of the modules back in.
CAUTION!
Be certain to plug the Blood Pressure
module
into the Blood Pressure or Colin position and not another module!
c)
IMPORTANT!
Turn the machine on (fan running)! If possible, return to Dialysis OR Heat Disinfect.
d) Re-m12 and / or +5 volts (within range now?). If the symptom was intermittent call debug
screen 1 to watch
5V Est
and / or
12V Est.
Allowing sufficient time, does the symptom you were
troubleshooting reoccur?
Yes
Symptom reoccurs! The last module plugged in (or its cable) is causing the problem!
No Screen on and/or good voltage(s)! Repeat parts a through d until ALL modules and have been
plugged in. If the symptom does not reoccur, reseating the cables MAY have eliminated the
problem.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...