![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635028.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 24
F- 1.0.4 FILACT = 1 OR CHAMBER #69 SAYS “AIR”
a)
FIGURE BELOW
, place one of the plugs, from the
FOUR-RESISTOR SET
, into the Air
Sensor’s distribution board
position i.e.
5
th
position
*
from the LEFT.
*
NOTE:
If CBE modified the resistor plugs into the CBE board pin for pin!
b) Based on Chamber #69’s text box now. TWO (2) possible scenarios:
1) IF (and ONLY if) it says “No Air” always:
See
(page 26).
2) IF says “Air” OR toggles between “Air” and “No Air”:
To avoid unnecessary work, perform
parts a THROUGH c below:
a)
BE CERTAIN
the resistor plug is placed PROPERLY in the
5
th
position
from the left!
b) If CBE modified, using a flashlight, ENSURE the resistor plugs into the CBE board, pin for pin.
ENSURE the CBE board’s TOP pin is covered by the resistor plug!
c) If (and ONLY if) the text box now says “No Air” always see
26). If it EVER says “Air” see
part d next page
!
Part d next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...