2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 538
BC LEAK- 3.0.0 DIAPHRAM LEAK TEST
Two (2) buckets are required, one empty and the other filled with one gallon of
ACID CONCENTRATE
. Place
both buckets on the floor and read the procedures before continuing:
a) Place the machine into RINSE.
If the external flow indicator’s ‘bob’ is not moving up and down
these tests are invalid!
b) Call debug screen 5. Allow
FPRE
(middle column, ABOVE
FPOS
), to become less than two thousand
eight hundred (2800). This verifies all conductive fluids have been rinsed out.
c) ENSURING RINSE is running (Remaining Time is NOT = 0:00) open the shunt door.
d) NOTE
FPOS
! Typically between 2000 and 4000.
e) Per the Figure right, direct the Ultrafiltrate Output Sample
Connector into the EMPTY BUCKET to prevent excessive
UF circuit pressure.
f)
IMPORTANT!
Per the Figure below, TIGHTLY clamp the
tubing at Valve #29’s. Ignore Flow Errors for now!
g) Remove the blue dialyzer connector from the shunt door and place it into the empty bucket.
h) Place the red dialyzer connector into the bucket of acid concentrate but
DO NOT
close the shunt door!
Rinse must NOT interrupt! The blue connector MUST be delivering flow into its bucket!
i)
WITHOUT LOOKING AWAY, watch
FPOS
for up to 45 seconds. Does it EVER, even if only once,
increase or fluctuate to more than seven thousand (7000)?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...