2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 399
EOS- 3.0.0 BICARBONATE (BIC) SYSTEM PROBLEM / ISOLATE THE BIC PUMP
This procedure uses the known good acid pump to check the bicarbonate pump.
A)
IMPORTANT! Turn the machine off to prevent damage!
B) Per the Figure below, unplug the bicarbonate pump’s cable from distribution board position “P17, BIC-P”.
CAUTION!
Avoid the Heparin Pump connector at position “P18”!
C) Plug the acid pump’s cable, into the bicarbonate pump’s position “P17, BIC-P”.
CAUTION!
The
connector should insert easily if the pins are okay!
D) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
E) Call debug screen 0. If debug does not appear ’Escape’ then call screen 0.
F) Allow thirty (30) seconds THEN,
ignoring the CONC symbol
, watch the BIC symbol for up to two
minutes:
1) EVER pink (solid or blinking)?
OR
2) Remaining ALWAYS white
?
1)
IF Pink:
The bicarbonate pump is probably okay. See procedure number
(page 400).
2) IF remaining White:
See
parts a THROUGH c below:
a)
IMPORTANT! Turn the machine OFF to prevent damage!
b) Return the acid pump’s cable to position “P16, CONC-P”.
b) Per the Figure (above), the bicarbonate pump is bad.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...