2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 109
F- 12.3.0 EITHER -250 MMHG COULD NOT BE ACHIEVED OR IT DID NO HOLD / ISOLATE FOUR-WAY
a)
Figure right
, clamp BOTH Four-Way dialyzer tubing
segments.
b) Can you achieve -250 mmHg and HOLD it
(+/- 15 mmHg) for one (1) minute now?
Yes
-250 can be achieved AND it held! See
procedure number
(page 109).
No -250 could NOT be achieved AND / OR could not be held! Either the transducer protector at the
meter is wet OR a Four-Way tubing connection is leaking. See parts A through C below:
A) Turn the machine OFF!
B) Locate and repair the leak.
C) Return to (ABOVE) procedure number
page 107
).
F- 12.3.2 -250 ACHIEVED AND HOLDS
a)
IMPORTANT!
Remove BOTH clamps from the Dialysate Line tubing segments.
b) Turn the machine OFF!
c) Turn the machine on and r
eturn to Dialysis Program (“Select Program” → ‘Dialysis’ →
‘
CONFIRM’)!
d)
IMPORTANT!
Turn Dialysate Flow OFF (Flow on/off lamp blinks).
e) Proceed to
page 466
, procedure number
LEFT BLANK INTENTIONALLY