2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 56
F- 6.1.6 ISOLATE DIALYSATE PRESSURE (PDIAL)
Call debug screen 4 to again watch
PDIA
for TWO (2) FULL minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) PDIA increases to more than 8.0 OR drops to less than 2.0:
Return to
page
2) IF PDIA NEVER increases to more than 8.0 OR drops to less than 2.0
: As NOTED in procedure
(page 50), drain flow was either:
1)
Pulse – Stop – Pulse – Stop – etc. (NO ‘dribble’)
OR; 2
)
Pulse – Stop – Pulse – ‘dribble’ (a noticeably weaker stream) etc. TWO (2) possible scenarios
i) or ii) below:
i)
IF (and ONLY if) drain flow = Pulse – Stop – Pulse – Stop (NO ‘dribble’)
:
See procedure
(page 56).
ii) IF drain flow = Pulse – Stop – Pulse – ‘Dribble’ (two second slow stream):
A
procedure, in a
different Section, is performed next.
NOTE
this page and procedure number (F- 6.1.6), because
you may prompted to return to here, THEN perform part A through C below:
A) BEFORE continuing to part b proceed to
page 535
, to perform
A LEAKING BALANCING CHAMBER DIAPHRAGM
B) If a balancing chamber diaphragm leak was not located in part a, return to Dialysis Program
(“Select Program”
→ ‘Dialysis’ →
‘
CONFIRM’).
C) See
part b
(page 56).
F- 6.1.7 ISOLATE FLOW PUMP CONTROL (FLWP)
a) If (and ONLY if) you have EVER seen [
Temperature
] ‘bounce’ up and down, more than 0.5º C, with
less than one (1) second in between, TWO (2) procedures: #
1)
NOTE
this page and procedure number
(F- 6.1.7) for as you may be prompted to return here; #
2)
BEFORE continuing to part b proceed to
page 535
SECTION 19 - TESTING FOR A LEAKING BALANCING CHAMBER DIAPHRAGM
b) Call debug screen 1.
FLWP
(upper right column) is a digital-to-analog value that controls Flow Pump
speed. It should remain stable (+/- 2). Without looking away watch
FLWP
for up to THREE (3) minutes.
Does it increase steadily more than 4, until a Flow Error occurs then falls only to increase again until a
Flow Error occurs, etc.?
Yes
FLWP
ramping up and down more than +/- 2: Possibly a tubing restriction somewhere. Start with
the Diasafe
®
filter possibly even inside the
housing.
No
FLWP
remaining stable: See procedure number
(page 57).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...