2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 79
F- 7.4.3 ISOLATE ACTUATOR-TEST BOARD
Per the Figure below, touch valve #30’s (black) solenoid. Is it HOT?
Yes
Valve #30’s solenoid is hot! See procedure number
(page 79).
No Valve #30’s solenoid is NOT hot! See parts a THROUGH c below:
a) Return valve #38 AND #31 to their distribution board positions.
b) Return the Deaeration AND Flow Pump to their distribution board positions.
c) NOTE
ONLY
VALVE #30 will be checked and proceed to
page 192
Figure 19 – Valve #30
F- 7.4.4 ISOLATE VALVE #30
a) Per the Figure above, check through valve #30’s INPUT and OUTPUT tubing for bio-growth restrictions.
b) Remove valve #30’s INPUT tubing and attach a syringe, with tubing attached that will fit snug over the
valve’s nozzle.
c) Push or pull on the syringe plunger. If valve #30 is open the plunger moves with no resistance. TWO (2)
possible scenarios:
1) IF the plunger offers NO resistance i.e. valve #30 open:
80).
2) IF the plunger offers resistance i.e. valve #30 restricted:
TWO (2) possible bad components:
1)
Valve #30 OR;
2)
Actuator-Test Board*.
*
To LOCATE the Actuator-Test board, refer to
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...