2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 361
CO- 6.8.0 CONDUCTIVITY IS NOT BETWEEN 13.0 AND 14.2
Based on the external meter’s reading, TWO (2) possible scenarios:
1)
IF (and ONLY if) LESS THAN 14.5 mS:
Proceed to
page 362
2) IF MORE THAN 14.5 mS:
See parts a THROUGH d below:
a) ENSURE the blue (bicarb) connector is not accidently plugged into acid; the red (acid) connector
MUST be in acid!
b) Figure below, clamp the tubing at Valve #29.
c) ENSURE the external flow indicator’s ‘bob’ continues to move up and down
d) Allowing up to fifteen (15) minutes, does the external meter’s reading stabilize between 13.0 and
14.4 (Yes or No)?
Yes
Conductivity between 13.0 and 14.4! Remove the clamp and allow (7) seven minutes.
If the meter’s reading becomes more than 14.5 mS again this confirms a problem with
Valve #29. TWO (2) possible bad components:
1)
Bad Actuator-Test Board OR;
2)
Bad Valve #29.
No
Conductivity is IS NOT between 13.0 and 14.4! See procedure number
(page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...