2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 57
F- 6.1.8 VERIFY FLOW RATE (BC SWITCH)
a)
Do NOT reset alarms!
b) Call debug screen 6. WITHOUT LOOKING AWAY, watch
BC Switch
(middle
column) for two (2) minutes OR until it EVER = 897 or more. FOUR (4) possible
scenarios 1) or 2) or 3) or 4) below:
1)
IF (and ONLY if) BC Switch REMAINS ALWAYS = 897:
ENSURING a “No Water” alarm has
NEVER occurred proceed to
page 89
, procedure number
2)
IF (and ONLY if) BC Switch ‘cycles’ to more than four-hundred (400) for two (2) seconds
then to less than 3.0 for nine (9) seconds:
Proceed to
page 62
, procedure number
3)
IF (and ONLY if) BC Switch ‘cycles’ to more than four-hundred (400) but NEVER to 897:
(page 57).
4)
IF BC Switch is NEVER more than four-hundred (400):
See parts a THROUGH c below::
a) Call debug screen 1. If
NPHT
(upper right) is decreasing press and hold
the “1” key until it is more than 600 then see part b If (and ONLY if)
NPHT
= 720 skip to part c.
b)
Allow forty-five (45) seconds BEFORE continuing to part c!
c) Return to debug screen 6 and continue to watch
BC Switch
for two (2) minutes. It may change
slightly but should
REMAIN ALWAYS
between 203 and 248?
Yes
BC Switch
is REMAINING between 203 and 248! The Flow Error is not
presenting at this time. It may be extremely intermittent! Return to
page 44
No
BC Switch
is NOT remaining between 203 and 248! Call the Home screen. TWO
(2) possible scenarios below:
1)
IF (and ONLY if) [Dialysate Flow] has remained at 800 ml/min:
See
2) IF [Dialysate Flow] has NOT remained 800 ml/min
: Set it to 800, allow thirty
(30) seconds, then return to (ABOVE) procedure number
page 57
).
F- 6.1.9 DETERMINE FLOW ERROR
From the Home screen, is [
Conductivity
] more than 14.8 mS?
Yes
[
Conductivity
] = More than 14.8 mS! Proceed to
page 62
procedure number
No
[
Conductivity
] is less than 14.8 mS! Proceed to
page 61
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...