2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 360
CO- 6.6.0 POST DIALYZER COND / TROUBLESHOOTING POST DIALYZER COND
a)
Recheck Loading Pressure (gauge in Rinse port). Unless there is an unannounced “No Water” or a
Flow Error alarm pressure cycles, to about 25 psi about every three (3) seconds! Do NOT continue
UNLESS cycling!
b)
ENSURING the shunt door is CLOSED, if the external flow indicator‘s ‘bob’ is NOT moving up and
down there may be an unannounced “No Water” or Flow Error. Do NOT continue UNLESS ‘bob’ is
moving up and down!
c)
Check for air bubbles flowing through the dialyzer AND
the dialyzer lines. Air does NOT allow conductivity to
stabilize!
d)
Figure right, using a flashlight, and WITHOUT
LOOKING AWAY for ONE (1) MINUTE watch for air
bubbles through the tubing at Cond Cell #7.
e)
DO NOT continue UNLESS ABSOLUTELY SURE
there are NO AIR leaks!
f)
See procedure number CO- 6.6.2 (page 360).
CO- 6.6.2 ISOLATE THE DIALYZER
a) Remove the dialyzer and reinstall the external meter to measure conductivity.
b) Close the shunt door then return to
Service Mode → Calibrate Sensors → Cond Cells.
c) Follow the screen’s instructions through:
“4. Waiting until the Conductivity value is stable.”
d) ENSURING the external flow indicator’s ‘bob’ is moving up and down, ALLOW six (6) FULL minutes to
ENSURE measured conductivity is between 13.0 and 14.4 mS.
e) Enter the measured conductivity into the
[Conductivity]
box and press ‘CONFIRM’ twice.
f) This time
DO NOT
install a dialyzer!
g) Press ‘CONFIRM’. Allow up to twenty-five (25) FULL minutes OR until if an alarm occurs. If all FOUR
‘OLC points’ advance successfully the screen says:
“Press ‘CONFIRM’ to complete the
calibration?”
Yes
“Press ’CONFIRM…”
appears! There may be a problem with the original dialyzer! ‘Escape’
the calibration. This time using a DIFFERENT PRIMED dialyzer return to
page 354
No
“Press ‘Confirm…”
DOES NOT appear! TWO (2) possible scenarios:
1)
IF (and ONLY if) “COND LESS THAN 10.0” OR “COND LESS THAN 15.0”
banner occurred:
(page 359).
2) All OTHER scenarios:
A possible problem with the Cond Cell #7 Circuit.
Proceed to
page 363
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...