2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 358
CO- 6.4.2 SECOND / THIRD / FOURTH OLC POINTS
a) Unless a “COND LESS THAN…” banner appears OR an announced Flow Error occurs. the remaining
three steps advance as below:
•
“9. Reading second point for post conductivity cell.”
Within
six (6) minutes
the
[Conductivity]
box should stabilize between 13.00 and 14.00 mS causing advancement to step #10 (third point)
•
“10. Reading third point for post conductivity cell.”
Within six (6) minutes
the
[Conductivity]
box should stabilize between 15.00 and 15.90 mS causing advancement to step #11 (fourth point)
•
“11. Reading fourth point for post conductivity cell.”
Within six (6) minutes the
[Conductivity]
box should stabilize between 13.00 and 14.00 mS
b) If all steps advance successfully the screen says:
“Press ‘CONFIRM’ to complete the calibration.”
?
Yes
“Press ‘CONFIRM’ to complete the calibration”
appears! See procedure number
(page 358).
No
“Press ‘CONFIRM’ to complete the calibration”
DOES NOT appear. TWO (2) possible
scenarios:
1)
IF (and ONLY if) “COND LESS THAN 10.0” OR “COND LESS THAN 15.0” banner
occurred:
Proceed to
page 359
, procedure number
2) ALL OTHER scenarios:
Proceed to
page 360
, procedure number
CO- 6.4.4 ALL FOUR OLC POINTS ADVANCED SUCESSFULLY
Press ‘CONFIRM’ to save the calibration. Does an “Error” banner occur?
Yes
Error banner!
Read before performing!
Acid clean the machine then return to (ABOVE)
procedure number
(page 354) but if you return here the Functional Board may be
bad (IC20)?
No
a) Turn the machine OFF.
b) Turn the machine on and return to Dialysis Program
(“Select Program” → ‘Dialysis’ →
‘CONFIRM’)!
c) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’
d) Allow six (6) minutes before continuing to part e!
e) If you were troubleshooting an OLC problem proceed to see
page 386
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...