2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 55
E) WITHOUT LOOKING AWAY, ENSURING a “No Water” banner NEVER appears, watch
BC
Switch
for up to five (5) minutes OR until if it EVER is more than 400 OR less than 203. FOUR
(4) possible scenarios 1) or 2) or 3) or 4):
1)
IF (and ONLY if) BC Switch is CONSTANTLY 897 or more:
Proceed to
page 89
2) IF (and ONLY if) BC Switch is NEVER more than 400 OR less than 203:
See
procedure
(page 55).
3)
IF (and ONLY if) BC Switch cycles CONSISTENTLY from about 220 for nine (9)
seconds to more than 400 (possibly 897) for two (2) seconds:
Proceed to
page 62
procedure number F- 7.0.0!
4) IF BC Switch is
EVER
more than 400 OR less than 203 but NOT CONSISTENTLY as
in Scenario #3:
Perform parts a through e below:
a)
Call debug screen 0. If Valve #24’s ‘dot’ is blue ENSURE the external flow indicator’s
‘bob’ is rising and falling.
b)
Turn Dialysate Flow OFF (Flow on/off lamp blinks).
c)
Figure right, if
ACFS
should return to between 4.0 and
6.0 instantaneously (within 0.5 seconds). If it EVER
increases slower than this OR remains less than 4.0 OR
more than 6.0 the CFS Transducer #10
*
is intermittent
bad.
*
To LOCATE the CFS refer to
(page 19
d)
Turn Dialysate Flow ON (Flow on/off lamp STOPS blinking).
e)
Repeat parts a through e at least eight (8) times. If
ACFS,
goes to between 4.0 and
5.9, within 0.5 seconds,
EVERY SINGLE TIME, see
F- 6.1.5 ACFS CYCLING / TMP ALARM
Call the Home screen. TWO (2) TMP window scenarios:
1)
IF (and ONLY if) the TMP window is white:
See
(page 56).
2) IF the TMP window is RED (TMP alarm):
See parts a THROUGH d below:
a) Press and release the ‘Reset’ key then immediately press and hold it for three (3) seconds.
b) Allow thirty (30) seconds.
c) If a TMP alarm reoccurs attempt RESET up to twice more BEFORE continuing to part d
d) Allow thirty (30) seconds BEFORE continuing to procedure number
(page 56).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...