2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 600
e) Per the Figure right AND Table 16 below,
TWO
(2)
measurements at the TEST Connector’s +5V AND +12V
pins. TWO (2) possible scenarios:
1)
IF (and ONLY if)
BOTH
are
IN RANGE
:
RECORD the +12V measurement then see procedure
number
(page 601).
2) ANY BAD measurement:
See
below and
respond accordingly.
Table 15 – Measurement Locations / Expected Good
Table 16 – Any BAD Measurement(s)
Me5
Volts (V
DC
)
Me12
Volts (V
DC
)
Response based on BOTH measurements
4.8 – 5.2
(i.e. good)
0 – 11.6
(i.e. low)
+5 volts GOOD; +12 volts LOW.
Proceed to
page
0 – 4.7
(i.e. low)
11.7 – 12.3
(i.e. good)
+5 volts LOW; +12 volts GOOD.
Proceed to
page
5.3 or more
(i.e. high)
11.7 – 12.3
(i.e. good)
+5 volts HIGH;
+12 volts GOOD. Proceed to
page 609
procedure number P- G.0.0
4.8 – 5.2
(i.e. good)
12.4 or more
(i.e. high)
+5 volts GOOD; +12 volts HIGH.
Proceed to
page 609
procedure number P- G.0.0.
5.3 or more
(i.e. high)
12.4 or more
(i.e. high)
BOTH +5 AND +12 volts HIGH! Proceed to
page 609
procedure number P- G.0.0.
0 – 4.7
(i.e. low)
0 – 11.6
(i.e. low)
BOTH +5 AND +12 volts LOW. Proceed to
page
, procedure number
Measurement
Pin # and location
Expected
(GOOD) range
+5V
2
(second pin from right)
4.8 – 5.2
V
DC
+12V
5
(five pins from right)
11.7 – 12.3
V
DC
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...