2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 296
Yes
LESS than 300
Ω!
(page 296).
No
MORE than 300 Ω! ENSURING the machine was OFF, prior to measuring, see parts a
THROUGH d:
a) ENSURE the
274 Ω
plug, from the
TWO-RESISTOR SET
, is in the
2
nd
position, from
the left, “MON-NTC”.
b) ENSURE the plug is aligned with the other connectors.
c) Use the
274 Ω
plug from a different
TWO-RESISTOR SET
d) Per the Figure below, measure again at pin 12 ENSURING good contact with the pin! Is the
meter’s numeric AND units display less than 300
Ω
now (Yes or No)?
Yes
LESS than 300
Ω!
(page 296).
No
MORE than 300 Ω! TWO (2) possible bad components: 1) Sensor cable OR
2) Distribution board.
T- 5.0.2
LESS THAN 300 Ω
A) Remove the
274 Ω
plug.
B) Return the
6.04 KΩ
plug, from the
TWO-RESISTOR SET
, to the
2
nd
distribution board position from the
left, “MON-NTC”.
C) Turn the machine on and r
eturn to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’).
(page 297).