2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 33
F- 2.6.0 MOTORS OKAY / ISOLATE ‘OPTIONAL’ DIALYSATE SAMPLER
a) Press the ‘Escape’ key.
b) At the bottom of the screen, select the Test & Options tab. DO NOT start the tests!
c) Per the Figure below, does the OPTIONAL
[Dialysate Sampler]
button
appear?
Yes The
[Dialysate Sampler]
button
should NOT appear! Refer to the NOTE below!
No
The
[Dialysate Sampler]
does NOT appear!
See
(page 33).
Test & Options Screen / Optional Dialysate Sampler button
F- 2.7.0 [DIALYSATE SAMPLER] DOES NOT APPEAR
Call the Home screen. Figure right, if the
TMP
window is RED a TMP alarm
is present. TWO (2) possible scenarios:
1)
IF (and ONLY if) the TMP window is WHITE:
(page 34).
2)
IF the TMP window is RED:
Perform parts a THROUGH d below:
a) Press and release the ‘Reset’ key then immediately press and hold it for three (3) seconds.
b)
Allow thirty (30) seconds!
c) If a TMP alarm reoccurs perform parts a and b up to twice more before continuing to part d.
d) Allow twenty (20) seconds then see procedure number
(page 34).
The
[Dialysate Sampler]
appears if the Dialysate Sampler
option has been accidentily
turned on.
FIGURE3C
This may
cause intermittent flow errors
and/or tubing to blow off. This
option MUST be set to “No” from
Servi
ce Mode → Options →
Hardware Options.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...