2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 366
CO- 8.0.0 MEASURED CONDUCTIVITY IS BAD / ‘ACTUAL’ CONDUCTIVITY CHECKS
TWO (2) possible scenarios based on the external meter’s reading versus TCD:
1)
IF (and ONLY if) LESS THAN TCD - (minus) 0.4 i.e. Conductivity is LOW:
Proceed to
page
, procedure number
2) IF MORE THAN TCD + 0.4 i.e. Conductivity is HIGH:
366).
CO- 8.1.0 CONDUCTIVITY IS HIGH (MORE THAN TCD + 0.4 mS)
TWO (2) possible scenarios based on the external meter’s reading:
1)
IF (and ONLY if) LESS THAN 15.0 mS:
(page 367).
2) IF 15.0 mS OR MORE:
See parts a THROUGH f below:
a) Turn the machine OFF!
b) Figure below, ENSURE the ACID AND BICARB PUMPS are not reverse connected at the
distribution board!
Continue to part c next page
c)
d)
e)