2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 367
f) Per the Figure below, trace the tubing from the red (ACID) connector to ENSURE it goes to the
ACID pump
g)
Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’).
h) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
i)
(page 367).
CO- 8.1.1 ISOLATE POSSIBLE VERY INTERMITTENT FLOW ERROR
a) Call debug screen 6.
b) WITHOUT LOOKING AWAY, watch
BC Switch
(middle column) for THREE (3) MINUTES If it EVER =
897 or more, even just once, indicates a masked Flow Error! Is it EVER = 897 or more?
Yes
BC Switch
= 897 even if just once! Proceed to
page 20
No
BC Switch
is NEVER, EVER = 897 or more! See procedure number
(page 368)