2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 62
F- 7.0.0 ISOLATE POSSIBLE RESTRICTIONS
a) ENSURE Dialysate Flow is ON (Dialysate Flow on/off lamp NOT blinking)!
b) Recheck Loading Pressure (Rinse port gauge). It may not cycle but MUST reach a PEAK between 23
and 27 psi?
Yes
Between 23 and 27 psi! See procedure number
(page 62).
No Less than 23 OR more than 27 psi! Return to
page 27
procedure number
F- 1.0.8
.
F- 7.0.1 LOADING PRESSURE OKAY (1)
A) Call debug screen 0. Watch
Flow Error
for one (1) minute. TWO (2) possible scenarios:
1)
IF Flow Error is ALWAYS = 1 (i.e. NEVER = 0):
2) IF Flow Error = 0 OR cycles between 1 and 0:
See part B.
B) ENSURING the machine has been in Dialysis Program at least five (5) minutes. Call the Home screen.
Is
[Conductivity]
= 17.0 mS?
Yes
[Conductivity]
= 17.0 mS! See procedure number
(page 63).
No
[Conductivity]
is less than 17 mS! Possibly even within range sometimes but then drifts
high. See part C
C) As bicarb precipitate may be partially plugging hydraulic pathways perform (2) two subsequent Acid
Cleans using vinegar or citric acid.
D) After the last Rinse cycle, return
to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
E) Allow [Conductivity] to increase to more than 13.0 BEFORE continuing!
F) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
G) Call debug screen 0.
H) WITHOUT LOOKING AWAY, watch
Flow Error
for up to FIVE (5) minutes OR until it EVER = 1, even if
just once! TWO (2) possible scenarios:
1)
IF (and ONLY if) Flow Error EVER = 1!
(page 63).
2) IF Flow Error ALWAYS = 0!
A flow error is not presenting at this time! Possibly the Acid Clean
fixed the problem!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...