2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 63
F- 7.0.2 LOADING PRESSURE ‘OKAY’ (2)
A)
To prevent damage, turn the machine OFF!
B) To avoid pulling cables loose, GENTLY open the card cage.
C) Per the Figure below, THREE (3) checks:
CHECK #1
CAREFULLY check the ENTIRE LENGTH of the Actuator Cable for damage!
CHECK #2
ENSURE the Actuator Cable is plugged in securely at both ends!
CHECK #3
ENSURE the valves are plugged into their CORRECT distribution board positions!
D) Was a problem located during the above three (3) checks?
Yes
ENSURING the problem was repaired, see procedure number
No problem found! Proceed to
page 64
, procedure number
F- 7.0.2.11 PROBLEM WITH ACTUATOR CABLE OR VALVE NOT PLUGGED IN WAS LOCATED
a)
Close the card cage!
b) Turn the machine on and r
eturn to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
c) Allow five (5) minutes then set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
d) Call debug screen 0, WITHOUT LOOKING AWAY, watch
Flow Error
for up to FIVE (5) minutes OR
until it EVER = 1, even if just once! TWO (2) possible scenarios:
1)
IF (and ONLY if) Flow Error EVER = 1!
(page 64).
2) IF Flow Error ALWAYS = 0!
A flow error is not presenting at this time!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...