2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 158
CLEAN- 1.2.0 NO LEAKS / VENT RESTRICTION?
NOTE:
Flow Errors may cause some Vent Tubing overflow. This is normal and can be ignored.
a) Per the
below, check for Vent Tubing kinks,
INCLUDING
the short segment going back to the
float!
b) If a kink was located and eliminated allow three (3) minutes to see if this fixes the flow error. If (and
ONLY if) the flow error reoccurs see procedure number
(page 159).