
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 562
LEAKING- 4.0.0 ISOLATE DRAIN FLOW (VALVE #30)
IMPORTANT!
A 60 ml luer lock syringe, FILLED with water is REQUIRED here!
a) Without causing tension on cables and tubing, move the hydraulics as far away from the cabinet as
possible!
b) ENSURE the distribution board cover is on!
c) Per Figure 107 below, check through Valve #30’s INPUT and OUTPUT tubing for restrictions.
d)
Turn the machine but when “Select Program” appears DO NOT press
‘Dialysis’ or start a Cleaning program!
e) Figure right, screw the 60 syringe onto the Fluid Sample Port.
f) Start a SIX (6) second timer, in your head, as you begin to
push AS
HARD AS YOU CAN
on the syringe plunger!
You should feel very little resistance!
g) Can you push ALL 60 ml, WITHOUT ‘blowing’ tubing off UF
Check Valve #64, within six (6) seconds?
Yes
All 60 ml can be pushed within 6 seconds! The Drain
path is okay! Proceed to
page 564
, procedure
No Cannot push all water out! See procedure number
(page 563).
Figure 107 – Hydraulics Top View
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...