2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 207
VEC- 1.1.0 BOB NOT MOVING / ISOLATE VALVE # 24
a) Call debug screen 2 (Figure below). If debug does not appear press ‘Escape’ then call screen 2.
b) Watch
DIAVLO
(lower left) for forty-five (45) seconds. BE CAREFUL, THREE (3) possible scenarios
1) or 2) or 3) below:
1)
IF (and ONLY if) DIAVLO cycles between 1 (for three seconds), and 0 (for one second):
(page 208).
2) IF (and ONLY if) DIAVLO ‘blinks rapidly’ to 0 but for way less than one second: NOTE
ONLY
VALVE #24
will be checked and proceed to
page 192
3)
IF DIALVLO REMAINS 0 OR 1 i.e. NEVER cycles:
See parts a THROUGH d below
a)
Turn the
machine OFF!
b) Figure right,
unplug Valve
#24’s from its
distribution
board position.
c)
Return to Heat Disinfect OR Rinse Program!
d) From debug screen 2, watch
DIALVLO
for forty five (45) seconds, does it now cycle
between 0 and 1?
Yes
Cycles between 0 and 1! Valve #24 or its blue wire harness is bad (to
LOCATE Valve #24 refer to
(page 193)).
No
Does NOT cycle!
A)
Return valve #24 to the distribution board;
B)
NOTE
ONLY
VALVE #24
will be checked and proceed to
page 192
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...