2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 208
VEC- 1.2.0 ‘BOB’ NOT MOVING / DIAVLO CYCLING BETWEEN 1 AND 0
a) Press ‘Escape’ to return to the Cleaning Program’s Main screen!
b) Leave the machine in whatever Program it is CURRENTLY in: 1) In Rinse? OR 2) In Heat Disinfect?
1) IF (and ONLY if) in RINSE:
2) IF in HEAT DISINFECT:
Per the Figure
right, does the VERY TOP Window,
Remaining Prerinse Time
= 0:00
min:sec (Yes or No)?
Yes
Remaining Prerinse Time
=
0:00! See procedure number
(page 208).
No
Remaining Prerinse Time
does
NOT = 0:00! See procedure number
page 209
VEC- 1.2.1 IN HEAT DISINFECT / REMAINING PRERINSE = 0:00
a) Call debug screen 0. If the screens do not appear press
‘Escape’ then call screen 0.
b) Only when
Remaining Prerinse Time
= 0:00, does Valve
#29’s ‘dot’ (Figure right) REMAIN blue (i.e. NEVER white).
ENSURE this by watching it for fifteen (15) FULL seconds.
c) Figure right, screw a 60 ml syringe, filled COMPLETELY
with water, onto the Fluid Sample Port.
d) Start a SEVEN (7) second timer in your head as you
begin to push
as HARD as you can
on the
syringe plunger! You should feel very little resistance!
e) Can you push ALL of the water through the Sample
Port within SEVEN seconds?
Yes Very little resistance is encountered! Valve
#29 is okay! See procedure number
VEC- 1.3.0 (page 209).
No
Significant resistance when pushing!
NOTE
ONLY
VALVE #29
will be checked and
proceed to
page 192
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...