2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 464
TMP- 3.4.0 ACFS NOT BETWEEN 3.5 AND 6.0 / ISOLATE PDIA
From debug screen 10, based on
PDIA
(middle column). TWO (2) possible scenarios:
1)
IF (and ONLY if) PDIA = less than 4.0:
(page 464).
2) IF PDIA = more than 6.0:
Proceed to
page 465
, procedure number
TMP- 3.5.0 PDIA LESS THAN 4.0 / ISOLATE TRANSDUCER #9
a) Figure right, unplug the Dialysate Pressure Transducer
#9 from distribution board position X9, “P-DIAL”.
b)
IMPORTANT!
To avoid unnecessary work,
using
the screen’s clock (upper right),
allow up to five (5)
minutes (
PDIA
response is NOT instantaneous).
PDIA
should increase to 9.0 or more?
Yes
PDIA
increased to 9.0 or more! See procedure number
(page 465).
No
PDIA
DOES NOT increase to 9.0 or more! See parts A through D below:
A) Turn the machine OFF.
B) Swap in the following components (
COMPONENT LIST
below), one at a time, with known
good, and in between perform parts C and D to see if the new component fixed this
problem.
COMPONENT LIST: 1)
Actuator-Test Board
2
;
2)
Sensor Board
1
;
3)
Sensor Board cable;
4)
Functional Board
1, 2
;
5)
Distribution board;
6)
Motherboard
1
1
To locate the boards refer to
(page 9)
2
To prevent “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not defined.
)
C) Leaving distribution board position “x9, “P-DIAL” VACANT return to Dialysis Program.
D) If screen 10’s
PDIA
is still NOT more than 9.0 return to part A!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...