2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 397
SECTION 6 – CONCENTRATE PUMP ERRORS
The CONC (Acid) and Bic (Bicarbonate) pumps each have optical End Of Stroke (EOS) sensors that
monitors the pump pistons when transported back and forth by their stepper motors. From debug screen 0,
per the Figure below, if all systems are okay, the
‘Actual’
number of steps (
AAcid
,
ABic
) REMAIN within
three (3) of the
‘Required’
number of steps (
Acid
,
Bic
). If not the pump symbol may turn pink.
EOS- 1.0.0 DETERMINE THE EOS ERROR
(page 398), ENSURE the pumps are installed with their output (red or blue) nozzles
towards the top and the cable extending from the side.
If not, this may be the problem!
B) Note this page number then perform
(page 6) and return here if a problem is NOT
located.
C) If the problem was occurring in a Cleaning Program (Heat Disinfect, Rinse, etc.) return to it. If the
problem was occurring in Dialysis Program:
1)
Return to Dialysis Program (
“Select Program” →
‘Dialysis’ → ‘CONFIRM’);
2)
From the Home screen, set [Dialysate Flow] to 800 ml/min!
D) Call debug screen 0. If debug does not appear press ‘Escape’ then call screen 0.
E) Allow thirty (30) seconds THEN, WITHOUT LOOKING AWAY, watch the pump symbols for up to two
(2) minutes OR until if one turns pink! FOUR (4) possible scenarios 1) or 2) or 3) or 4) below
:
1) IF the CONC (acid)
symbol is EVER pink (solid or blinking) AND the BIC
symbol
is ALWAYS white:
ACID pump system problem! See procedure number
(page 398).
2) IF the CONC (acid)
symbol is ALWAYS white AND the BIC symbol
is EVER
p
ink
(solid or blinking):
BIC pump system problem!. See procedure number
page 399
).
3) IF BOTH the CONC (acid)
AND the BIC
symbols are EVER pink (solid or blinking):
page 402
4) IF BOTH pump symbols REMAIN ALWAYS white (NEVER pink):
See parts a AND b below:
a) WITHOUT LOOKING AWAY, watch
AAcid
versus
Acid
for up to five (5) minutes. If
AAcid
REMAINS /- 3 of
Acid
the acid pump system is okay (for now anyway). In this event
continue to part b. If
AAcid
does NOT remain /- 3 of
Acid
then TWO (2) possible bad
components:
1)
Bad ACID pump OR;
2)
Bad Actuator-Test Board.
b) Watch
ABic
versus
Bic
for up to five (5) minutes. If
ABic
REMAINS /- 3 of
Bic
the
bicarb pump system is okay (for now anyway)! If
ABic
does NOT remaining /- 3 of
Bic
then TWO (2) possible bad components:
1)
Bad BICARB pump OR;
2)
Bad Actuator-Test
Board.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...