2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 282
T- 2.6.8 HEATER COMPONENTS SO FAR OKAY
a) If troubles
hooting a Dialysis Program problem return to DIALYSIS PROGRAM (“Select Program” →
‘Dialysis’ → ‘CONFIRM’). If a Heat Disinfect problem return to HEAT DISINFECT
b)
Set your volt meter to AC voltage (~ V, V
AC
)!
c) Figure right, at the distribution board’s Heater
Connector, measure again between the BLUE and
BROWN wires! More than 100 volts AC now?
Yes
More than 100 volts! The heater is on! See
(ABOVE) procedure number
(page 266).
No THREE (3) possible bad components (see
Component List below). With the machine off swap
them in, one at a time, in between repeating
procedure number
(page 282) to test the new
component until the heater is on!
Component List
:
1)
20-pin (X2) Power Logic cable;
2)
Triac (see Figure below);
3)
Power
Control board (inside the power supply)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...