2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 635
P- 1.2.0 PIN 4 LESS THAN 10 VOLTS / ISOLATE MAIN LINE FUSES 6.3 AMP FUSES
a)
Unplug the machine
!
CAUTION!
Electrocution hazard if not unplugged!
b)
IMPORTANT! Set your CALIBRATED
volt meter to RESISTANCE (Ω)!
c) To determine the meter’s resistance touch its leads together. The meter MUST read less than 1.0
Ω
!
Subtract this reading from all subsequent measurements!
d) On the Power Control board, per the
below
, measure
ACROSS
BOTH
6.3 Amp fuses.
NOTE: “OL” = bad fuse!
Figure 127 – Power Control Board / Fuses S12 and S16 / ST2 and ST1
e) Are BOTH fuses less than 1.0
Ω
?
Yes
BOTH less than 1.0
No One or both fuses MORE THAN 1
.0 Ω or “OL”! Proceed to
page 637,
procedure number
P- 1.2.1 BOTH FUSES LESS THAN 1.0
Ω / ISOLATE INCOMING (LINE) VOLTAGE
a) Plug the machine in
.
Caution!
High Voltage now present!
b)
IMPORTANT!
ENSURE the Power Supply’s Main Power Switch is
ON
(i.e. rocker switch pushed into
the “1” position)!
Parts c through e next page