SECTION 11 – INDUCED AIR LEAK TESTS
These procedures use negative pressure to locate an air leak. Use a flashlight to look THROUGH tubing. If
air is seen continue against the direction of flow (Figure below), to the next component until air bubbles are
not seen.
NOTE:
Air is NORMAL in some locations, for example through the input tubing of Deaeration Pump #20
AND in the ‘To Drain’ tubing.
(page 491), look through the Flow Pump’s (clear) input tubing for air bubbles that
START in the middle of the tubing segment.
If bubbles are located this may be the problem!
B)
Temperature
and
Conductivity
MUST remain normal and the external flow indicator’s ‘bob’ MUST be
rising and falling!
C) RESET ALL alarms!
D) If not already, run the alarms and pressure tests. At this point it does not matter if they pass!
E) Turn the blood pump on so that is is rotating at least 100 ml/min.
F) From the Home screen, set
[UF Goal]
to 1000; set
[UF Time]
to 1:00 Hr.
G) Press ‘CONFIRM’.
(page 490).
AIR- 1.0.0 ACHIEVE NEGATIVE DIALYSATE PRESSURE
This procedure describes how to acheive negative pressure (TMP). TMP MUST be maintained between 300
and 360 mmHg for all subsequent tests. If it EVER falls below 220 repeat parts a through c below:
a) RESET ALL alarms!
b) To turn UF on, Figure right, press the front panel’s UF on/off
key. The UF lamp turns on solid.
c) Figure right, allow the ‘actual’
TMP
to increase to 250 then
IMMEDIATELY
turn UF off
(UF Lamp off)!
(page 491).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...