2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 499
Loading Pressure less than 16 psi continued:
c) Return to Dialysis Program (“Select Progr
am” →‘Dialysis’ → ‘CONFIRM’)!
d) Proceed to
page 116
, procedure number
DAIR- 2.0.0 LOADING PRESSURE 16 PSI OR MORE
a) This is NOT a routine calibration! Follow the instructions to avoid error!
NOTES:
1) Loading and Deaeration pressure are calibrated until BOTH are in range. LOADING
between 23 and 25 psi; DEAERATION refer to
(page 667).
2) The value in the
[Pump Rate]
data box determines the speed of the Deaeration Motor
and thus Deaeration pressure!
[Pump Rate]
should NEVER be set to less than 180!
3) The BYPASS valve, on the Deaeration Pump (see Figure 84 (page 500)), should be
adjusted ONLY if Deaeration pressure is too high when
[Pump Rate]
= 180.
4) Loading Pressure is mechanically calibrated using Valve #65 (see Figure 84 (page 500))!
5) ‘Sharply’
press ‘CONFIRM’ after each
[Pump Rate]
adjustment. The data box MUST be
pale yellow/white (NOT bright yellow or gray!) for the rate adjustment to take effect.
b) Select the
[PUMP RATE]
data box, it turns bright yellow.
c) Adjust its value, then sharply press ‘CONFIRM’ ONCE. The box MUST return to pale yellow / white i.e.
not bright yellow or gray!
d) Repeat parts b and c until if target pressures can be achieved?
Yes
Target pressures achieved!
A)
Press ‘CONFIRM’ to save the calibration!
B)
Turn the
machine off then on and return to Dialysis Prog
ram (“Select Program” → ‘Dialysis’ →
‘CONFIRM’);
C)
The Deaeration problem is solved but return to the procedure that prompted
you here!
No Target pressures CANNOT be achieved! See procedure number
(page 500)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...