2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 308
T- 7.4.0 ‘BOB’ IS RISING AND FALLING / CHECK ‘ACUAL’ TEMPERATURE
Based on the external meter’s reading, THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) remaining less than 33.0° C:
2)
IF (and ONLY if) remaining between 33.1 and 38.9° C:
Proceed to
page 309
, procedure
number
3)
IF EVER WAS more than 38.9° C, even if only ONCE:
Perform parts a through c below:
a) Return to Dialysis Program (
“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
b) Using a temperature meter, ENSURE incoming water temperature is less than 35.0º C.
c) If (and ONLY if) incoming water temperature is less than 35.0° C proceed to
page 241
T- 7.4.1 TEMPERATURE LESS THAN 33° C
a)
To avoid damage turn the machine OFF!
b)
Read before performing!
Reseat the Power Logic, Sensor and Functional Boards then return to
page 305
). If you return here continue to part c..
c)
Read before performing!
Return to (ABOVE) procedure number
page 305
) but THIS TIME
set
[TEMP DAC]
to “180”. If you return here see (ABOVE) procedure number
(
page 257
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...