2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 257
T- 1.7.0 TEMPERATURE IS LOW / ISOLATE TEMPERATURE MONITOR CIRCUIT
a) Return the concentrate connectors to their rinse ports
b)
Place the machine in HEAT DISINFECT!
c) Being aware that flow does not stop, remove the dialyzer connectors from the shunt and connect them to
a Temperature (° C) meter but
DO NOT
shut the door
d) Ignoring a possible temporary “Flow Recirc Error” the “Cover is Open” banner presents.
e) ENSURING a “No Water” alarm NEVER occurs, the external flow indicator’s ‘bob’ MUST be moving up
and down!
(page 257).
T- 1.7.1 ISOLATE NTC #2
a) Figure right, unplug the connector from the
1
st
distribution
board position from the left, “CON-NTC”. This is NTC #2 and
unplugging it should turn the heater on.
b)
To avoid damage, DO NOT leave the machine!
c) The meter’s temperature should eventually increase. If incoming water is really cold response may take
up to seven (7) minutes. If incoming water is at normal temperature usually within three (3) minutes!
d) WITHOUT LOOKING AWAY, watch
the external meter for UP TO seven (7) minutes
OR
until if it
reaches 36.0° C OR more?
Yes
36.0° C OR more!
(page 257).
No Remains less than 36.0° C! Leaving NTC #2 unplugged, proceed to
page 258
T- 1.7.4 TEMPERATURE MORE THAN 36.0° C / ISOLATE MONITOR (NTC #3)
a) Call debug screen 4. If debug does not appear press ‘Escape’ then call screen 4.
b) From debug screen 4, is
TEMP
(lower left)
4.0 OR more?
Yes
TEMP
4.0 or more. Proceed to
page 259
, procedure number
No
TEMP
is less than 4.0! See parts THROUGH c below:
b) Replace NTC #3
*
with a known good.
*
To LOCATE NTC #3 refer to see
c) Proceed to
page 302
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...