T- 7.5.3 TEMPERATURE CHANGES MORE THAN +/- 0.2° C / NTC #2 CHECK
THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) NTC #2 was NEITHER checked NOR replaced in THIS troubleshooting session:
See procedure number T- 8.0.0 (page 315) to check (or replace) NTC #2
*
.
*
To LOCATE NTC #2 refer to
(page 218)
2) IF (and ONLY if) NTC #2 was checked in THIS troubleshooting session:
Replace NTC #2
*
with a
known good then return to (ABOVE) procedure number
(page 305) to see if this was the
problem.
*
To LOCATE NTC #2 see
(page 218)
3)
IF NTC #2 was replaced in THIS troubleshooting session: READ before continuing!
Swap in
the listed components (see COMPONENT LIST), one at a time, and in between return to
(ABOVE) procedure number
(page 305) to test each new component.
COMPONENT LIST: 1)
Sensor Board
*
;
2)
Power Logic Board
*;
3)
Heater
**
;
4)
Functional
Board
*
;
5)
TWO (2) power supply components:
i)
Power Control Board (inside the power supply);
ii)
Heater Triac***.
*
To LOCATE these boards refer to
(page 9)
**
(page 129)
***
To LOCATE the triac, refer to
(page 291).
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...