CO- 8.4.0 ISOLATE ACID PUMP
a) Turn the machine OFF!
b) Figure right, ENSURE the Acid (red) AND Bicarbonate
(blue) Connector’s O-rings AND ENSURE the filters are
clean AND the adaptor plugs are screwed in tight
c) Read parts d and e before performing them!
d) Perform the ACID (CONCENTRATE) PUMP VOLUME CALIBRATION per the Calibration booklet.
e) Are you able to enter measured burette volumes
AND ‘CONFIRM’ them without an “Operator Error”
(Figure right) occurring?
Yes
“Operator Error” did NOT occur! See procedure number
(page 370).
No “Operator Error” occurred! Acid pump problem! See (ABOVE) procedure number
CO- 8.5.2 ISOLATE BICARBONATE PUMP
a) Perform the BICARBONATE PUMP VOLUME CALIBRATION.
b) Are you able to enter measured burette volumes AND ‘CONFIRM’ them without an “Operator Error”
occurring?
Yes
“Operator Error” DID NOT occur! See procedure number
(page 370).
No “Operator Error” occurred! Bicarb Pump problem! See (ABOVE) procedure number
CO- 8.5.4 VERIFY CONDUCTIVITY
a) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → “CONFIRM’)!
b) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’!
c) Allow fifteen (15) minutes BEFORE continuing to part d.
d) ENSURING the external flow indicators’ ‘bob’ is moving up and down, is the external meter now within
+/- 0.4 mS of TCD?
Yes
Conductivity within 0.4 of TCD! The pump calibrations fixed the problem!
No Conductivity is IS NOT within 0.4 of TCD! TWO (2) possible scenarios:
1)
IF (and ONLY if) measured conductivity is LESS THAN TCD - (minus) 0.4 (i.e.
Cond is Low):
Proceed to
page 372
, procedure number
2) IF measured conductivity is MORE THAN TCD + 0.4 (i.e. Cond is High):
See
(page 371).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...